Evaluasi Diagnostic Reference Level dan Risiko Kanker pada Pemeriksaan CT Toraks

Nurpuspasari, Maytalia (2023) Evaluasi Diagnostic Reference Level dan Risiko Kanker pada Pemeriksaan CT Toraks. Other thesis, Institut Teknologi Sepuluh Nopember.

[thumbnail of 01111940000033-Undergraduate_Thesis.pdf] Text
01111940000033-Undergraduate_Thesis.pdf - Accepted Version
Restricted to Repository staff only until 1 October 2025.

Download (1MB) | Request a copy

Abstract

Pada bidang radiodiagnostik, terdapat upaya untuk membuat dosis yang diterima oleh pasien serendah mungkin dengan tetap menjaga kualitas citra seoptimal mungkin yang disebut optimisasi. Optimisasi dapat dilakukan dengan cara memantau dosis yang diterima oleh pasien. Salah satu kegunaan optimisasi ialah penyusunan Diagnostic Reference Level (DRL). Dosis radiasi yang diterima oleh pasien dapat menimbulkan perubahan pada organ atau jaringan dan dapat menimbulkan risiko kanker. Oleh karena itu, diperlukan adanya perkiraan risiko kanker akibat paparan radiasi. Penelitian ini akan membahas mengenai evaluasi nilai DRL dan risiko kanker pada pemeriksaan CT Toraks agar dapat mengurangi risiko kanker akibat radiasi pada pasien remaja hingga dewasa. Metode yang digunakan dalam penelitian ini yaitu mengumpulkan data sekunder dosis radiasi pasien Rumah Sakit Simpang Lima Gumul Kediri yang melakukan pemeriksaan CT Toraks. Setelah itu akan ditentukan nilai DRL berdasarkan nilai SSDE. Selain itu, ditentukan dosis efektif dan dosis organ setiap pasien untuk memperkirakan risiko kanker akibat paparan radiasi berdasarkan ICRP 103. Hasil DRL yang diperoleh yakni pada kategori remaja, dewasa, lansia, dan manula secara berturut-turut adalah 30,13 mGy, 29,73 mGy, 29,23 mGy, dan 28,66 mGy. Nilai SSDE yang berada di atas nilai DRL akan dievaluasi berdasarkan parameter penyinarannya. Nilai dosis efektif seluruh pasien masih di bawah Nilai Batas Dosis (NBD) yang ditetapkan oleh BAPETEN yaitu 50 mSv. Rata-rata dosis organ tertinggi pada CT toraks adalah kelenjar timus yaitu sebesar 41,3 mGy. Nilai dosis organ akan menurun seiring dengan bertambahnya usia. Faktor risiko kanker pada pemeriksaan CT Scan toraks pada kategori usia remaja 0,1 – 0,14%. Pada kategori usia dewasa berkisar 0,09 – 0,12%. Pada kategori usia lansia berkisar 0,09 – 0,14%. Dan kategori usia manula berkisar 0,09 – 0,13%.
===============================================================================================================================
In the field of radiodiagnostics, there is an effort to make the dose received by the patient as low as possible while maintaining the optimal image quality which is called optimization. Optimization can be done by monitoring the dose received by the patient. One of the uses of optimization is the preparation of a Diagnostic Reference Level (DRL). The radiation dose received by the patient can cause changes in organs or tissues and can increase the risk of cancer. Therefore, it is necessary to estimate the risk of cancer due to radiation exposure. This study discuss the evaluation of DRL values and cancer risk on chest CT examinations in order to reduce the risk of radiation-induced cancer in adolescent to adult patients. The method used in this study was to collect secondary data on radiation dose of patients at Simpang Lima Gumul Hospital, Kediri who performed a CT chest examination. After that, the DRL value determined based on the SSDE value. In addition, the effective dose and organ dose for each patient were determined to estimate the risk of cancer due to radiation exposure based on ICRP 103. The DRL results obtained were in the adolescent, adult, younger elderly and older elderly categories respectively 30.13 mGy, 29.73 mGy, 29.23 mGy, and 28.66 mGy. The SSDE value which is above the DRL value will be evaluated based on the irradiation parameter. The effective dose value for all patients is still below the Dose Limit Value set by BAPETEN, which is 50 mSv. The highest organ dose on chest CT was the thymus, which was 41.3 mGy. Organ dose values will decrease with age. Cancer risk factors on chest CT scanning in the adolescent age category range from 0.1 – 0.14%. In the adult age category it ranges from 0.09 – 0.12%. In the younger elderly age category it ranges from 0.09 – 0.14%. And the older elderly age category ranges from 0.09 – 0.13%.

Item Type: Thesis (Other)
Uncontrolled Keywords: CT-Scan, Dosis Efektif, DRL, SSDE, Toraks, Effective Dose, Thorax
Subjects: Q Science > QC Physics
R Medicine > R Medicine (General)
Divisions: Faculty of Science and Data Analytics (SCIENTICS) > Physics > 45201-(S1) Undergraduate Thesis
Depositing User: Maytalia Nurpuspasari
Date Deposited: 24 Aug 2023 08:54
Last Modified: 24 Aug 2023 08:54
URI: http://repository.its.ac.id/id/eprint/102517