Desain dan simulasi komponen sumber tegangan tinggi pada pembangkit sinar-x untuk peralatan diagnostik medis

Wiranata, Addien Wahyu (2015) Desain dan simulasi komponen sumber tegangan tinggi pada pembangkit sinar-x untuk peralatan diagnostik medis. Undergraduate thesis, Institut Teknologi Sepuluh Nopember.

[thumbnail of 2211100071-Undergraduate_Thesis.pdf]
Preview
Text
2211100071-Undergraduate_Thesis.pdf

Download (6MB) | Preview

Abstract

Pembangkit sinar-x pada peralatan diagnostik medis membutuhkan
tegangan kerja sebesar 50 kV- 150 kV berdasarkan kebutuhan daya
tembus yang disesuaikan dengan intensitas volume tubuh pasien.
Tegangan kerja tersebut dihasilkan dengan menggunakan transformator
berdaya tinggi yang memiliki ukuran sangat besar. Sehingga
transformator tersebut membutuhkan ruangan besar yang disertai
mahalnya harga dari transformator berdaya tinggi dan sulitnya
didapatkan di pasaran lokal. Untuk itu pada penulisan tugas akhir ini
akan dilakukan penelitian dengan menggunakan rangkaian elektronika
daya yang didesain untuk menghasilkan frekuensi maksimal 15 kHz
sebagai faktor variabel yang membangkitkan tegangan kerja pada
peralatan diagnostik medis. Rangkaian elektronika daya ini
menggunakan intermediet circuit dan inverter satu fasa. Intermediet
circuit digunakan unuk menaikkan tegangan yang sebelumnya telah
disearahkan dari sumber PLN sebesar 220 volt. Namun dalam
penelitian tugas akhir ini tidak dihilangkan komponen utama
transformator yang akan dianalisa sebagai nilai histerisis dari high
frequency losses transformator. Sehingga tegangan yang dihasilkan
diharapkan mencapai nilai rentang 50 kV-150 kV dengan frekuensi
maksimal 15 kHz.

===============================================================================================

Generating X-Rays in medical diagnostic equipment requires a
working voltage of 50 kV kV- 150 by penetrating power requirements
adapted to the intensity of the volume of the patient's body. The working
voltage is generated by using a high-power transformer which has a very
large size. So that the transformer needs a large room with the high price
of high-power transformer and the difficulty found in the local market.
For it was on this thesis research will be conducted using power
electronics circuits that are designed for a maximum frequency of 15
kHz produces a voltage variable factors that generate employment in
medical diagnostic equipment. This power electronic circuit using the
intermediate circuit and an one phase inverter. Intermediates used unuk
boost circuit that previously had been rectified from PLN source of 220
volts. But in this research is not eliminated the transformer main
components will be analyzed as hysteresis values of high frequency
transformer losses. So that the voltage generated is expected to reach a
value range of 50 kV-150 kV with a maximum frequency of 15 kHz.

Item Type: Thesis (Undergraduate)
Additional Information: RSE 621.314 Wir d
Uncontrolled Keywords: peralatan diagnostik medis; intermediet circuit; inverter satu fasa; high frequency losses transformator
Subjects: T Technology > TK Electrical engineering. Electronics Nuclear engineering > TK6565.T7 Transformers
Divisions: Faculty of Electrical Technology > Electrical Engineering > 20201-(S1) Undergraduate Thesis
Depositing User: - Taufiq Rahmanu
Date Deposited: 09 Oct 2019 07:12
Last Modified: 09 Oct 2019 07:12
URI: http://repository.its.ac.id/id/eprint/71082

Actions (login required)

View Item View Item