Tingginya angka kematian di Indonesia disebabkan oleh kecelakaan kendaraan
bermotor. Menurut data statistik, pada tahun 2015 kecelakaan pengendara motor
mencapai 14 kali dalam sehari sehingga berakibat pada banyaknya korban yang
mengalami patah tulang. Tindakan medis untuk mengatasi permasalahan tersebut adalah
dengan menyambung tulang yang patah menggunakan bone plate dan dipasangkan
dengan screw. Oleh karena itu perlu dilakukan penelitian tentang implant ortopedik
untuk kasus patah tulang yaitu dengan pembuatan bone plate yang sesuai dengan ukuran
tulang orang Indonesia. Desain bone plate pada penelitian ini dikerjakan dengan
software ANSYS 15.0 menggunakan parameter geometri pada lubang plate, serta
pemilihan material yaitu SS304 dan SS316L. Desain bone plate meliputi plate duo,
flower dan sloted yang dipasangkan dengan screw diameter 4,5 mm, 4,0 mm, 3,5 mm,
3,0 mm, dan 2,7 mm masing masing dengan posisi 1 sampai posisi 6. Pemberian beban
sebesar 550 N diterapkan pada masing masing desain. Hasil yang diperoleh adalah
tegangan, regangan, dan deformasi dari plate, screw dan tulang. Kombinasi plate duo
dengan diameter screw 4,5 mm didapatkan hasil desain optimum pada posisi 1 dan 2
untuk material SS304 karena nilai tegangan yang dihasilkan berada dibawah tegangan
luluh dari material SS304. Plate flower mengalami kegagalan disemua desain karena
nilai tegangannya berada diatas tegangan luluh material dan plate sloted didapatkan
hasil yang optimum pada posisi 1 sampai 5 untuk material 316L sedangkan untuk
material 304 memiliki desain terbaik pada semua posisi. Kombinasi plate sloted
merupakan desain terbaik dari yang lainnya karena secara keseluruhan tegangan yang
dihasilkan berada dibawah tegangan luluh materialnya
=====================================================================================================
High mortality rate in Indonesia is caused by motor vehicle accidents.
According to the statistics in 2015, motorcycle crash hit 14 times during a day and cause
many victims suffere the broken bone. This problem is resolved by medical measure
which is connects the fractured bone with a bone plate and paired a screw. Therefore,
research on orthopedic implants for fracture is done by making a bone plate suitable with
the size of the bone of Indonesian people. Design bone plate in this research is done by
ANSYS 15.0 software using geometric parameters on the hole plate and material
selection is SS304 and SS316L. Design bone plate cover plate duo, flower and sloted are
paired by a screw diameter of 4.5 mm, 4.0 mm, 3.5 mm, 3.0 mm and 2.7 mm respectively
with position 1 to position 6. Giving load of 550 N is applied to each design. The results
are obtained stress, strain, and deformation of the plate, screw and bone. Combination of
plate duo with a screw diameter of 4.5 mm are obtained optimum design at positions 1
and 2 for SS304 material because the value of stress is below the yield strength of the
SS304. Combination of plate flower failures in all designs because the value of stress is
above the yield strength of the material and plate sloted is optimum result at positions 1
to 5 of material 316L while the material 304 has the best design in all positions.
Combination of plate sloted is the best design from the others because value of stress is
below the yield strength of the material

Item Type: Thesis (Masters)
Additional Information: RTMt 620.112 23 Shi p
Uncontrolled Keywords: ANSYS, Beban, Bone Plate, Geometri.
Subjects: T Technology > TA Engineering (General). Civil engineering (General) > TA418.74 Corrosion and anti-corrosives
Divisions: Faculty of Industrial Technology and Systems Engineering (INDSYS) > Material & Metallurgical Engineering > 27101-(S2) Master Thesis
Depositing User: Mr. Tondo Indra Nyata
Date Deposited: 20 Mar 2020 06:55
Last Modified: 20 Mar 2020 06:55
URI: http://repository.its.ac.id/id/eprint/75520

Actions (login required)

View Item View Item