Rancang Bangun Sistem Monitoring Laju Aliran Dengan Metode Perbedaan Tekanan Pada Pitot Tube Menggunakan Sensor Mpx2010dp Berbasis Arduino

A’imah, Hilyatul (2016) Rancang Bangun Sistem Monitoring Laju Aliran Dengan Metode Perbedaan Tekanan Pada Pitot Tube Menggunakan Sensor Mpx2010dp Berbasis Arduino. Diploma thesis, Institut Teknologi Sepuluh Nopember Surabaya.

[thumbnail of 2413031070-Non_Degree.pdf]
Preview
Text
2413031070-Non_Degree.pdf - Published Version

Download (2MB) | Preview

Abstract

Dalam sistem pengukuran aliran, terdapat beberapa metode yang digunakan salah satunya adalah metode perbedaan tekanan atau head flow meter. Prinsip kerja dari head flow meter adalah dengan memasang penghalang yang menyebabkan tekanan dan alirannya berubah. Dengan mengukur perbedaan tekanan antara sebelum dan sesudah penghalang maka dapat ditentukan besarnya aliran fluida dalam pipa tersebut. Pitot tube merupakan salah satu alat ukur yang menggunakan metode perbedaan tekanan. Terdapat dua lubang pengukuran yaitu tekanan stagnasi dan tekanan statis. Setiap alat ukur memiliki spesifikasi dan karakteristik yang mempengaruhi hasil pengukuran, sehingga perlu dilakukan pengukuran dan kalibrasi pada alat ukur untuk mencari karakteristik suatu alat ukur. Pada penelitian kali ini telah dibuat sebuah rancang bangun sistem pengukuran dan monitoring laju aliran pada pitot tube dengan sensor MPX2010DP yang diintegrasikan dengan arduino uno sebagai mikrokontrollernya dan visual basic 2013 sebagai interface dari monitoring. Alat ukur dari rancang bangun ini memiliki nilai Cv sebesar 18,218x , dimana setiap alat ukur memiliki nilai Cv yang berbeda-beda tergantung dari bentuknya. Berdasarkan hasil pengujian alat didapatkan nilai ketidakpastian ±0,0546 dengan tingkat kepercayaan 95%. Sedangkan nilai keakurasiannya mencapai 98,55% dengan error 0,0145
============================================================================================
In the flow measuring system, one of a methods for measuring flow is head flow meter. Basic principle of work of head flow meter is installing the barriers which cause the pressure and its flow change. by measuring pressure difference between the barriers before and after installed, the fluids flow can be determined. Pitot tube is one of device for measuring system that uses pressure difference methods. The stagnation pressure and static pressure are the two mesurement holes. Every mesurement device has certain specification dan properties that influence the measurement result, so the measurement device needs to calibrate in order to find the characteristic of the device. In this experiment, the design of flow rate measuring and monitoring system in the pitot tube with MPX2010DP sensors that is integrated with arduino uno as its microcontroller dan Visual Basic 2013 as its interface from monitoring has been made. The measurement device from this design has Cv value of 18,218x, which every measurement device has its own different Cv value depends on its shape. Based on testing result, the uncertainty value of device is ± 0,0546 with reliability of 95%. Whereas the accuracy value is 98,55 % with 0,0145 error

Item Type: Thesis (Diploma)
Additional Information: RSF 621.391 6 Aim r
Uncontrolled Keywords: Laju aliran, pitot tube, monitoring
Subjects: T Technology > TA Engineering (General). Civil engineering (General) > TA1573 Detectors. Sensors
T Technology > TJ Mechanical engineering and machinery > TJ935 Pipe--Fluid dynamics. Tubes--Fluid dynamics
Divisions: Faculty of Industrial Technology > Physics Engineering > (D3) Diploma 3
Depositing User: EKO BUDI RAHARJO
Date Deposited: 30 Jul 2020 05:23
Last Modified: 30 Jul 2020 05:24
URI: http://repository.its.ac.id/id/eprint/76724