Rancang Bangun Sistem Monitoring Laju Aliran Pada Elbow Tabs Dengan Menggunakan Sensor MPX2010DP Berbasis Arduino

Suwartika, Yurike (2016) Rancang Bangun Sistem Monitoring Laju Aliran Pada Elbow Tabs Dengan Menggunakan Sensor MPX2010DP Berbasis Arduino. Diploma thesis, Institut Teknologi Sepuluh Nopember.

[thumbnail of 2413031078-Non_Degreee.pdf]
Preview
Text
2413031078-Non_Degreee.pdf - Published Version

Download (2MB) | Preview

Abstract

Pengukuran laju aliran fluida merupakan pengukuran untuk mendapatkan data-data untuk dianalisis karena pada prinsipnya fluida merupakan energy bertekanan yang semakin banyak dikembangkan pada proses-proses di industri. Banyak metode pengukuran laju aliran yang digunakan dalam mekanika fluida, setiap metode yang digunakan akan menghasilkan suatu nilai karakteristik yang berbeda-beda, tidak semua jenis pengukuran laju aliran dapat menggunakan metode yang sama, dilihat dari pressure drop yang dihasilkan akan mempengaruhi setiap metode yang digunakan. Pada penelitian tugas akhir ini dilakukan pengukuran laju aliran dengan menggunakan metode elbow tabs. Pada pengukuran fluida sering dijumpai sambungan pipa seperti elbow dengan kemiringan tertentu seperti 45° atau 90°. Kemiringan pada elbow tersebut dapat mempengaruhi perbedaan tekanan maupun pressure drop pada fluida yang melewati elbow tersebut. Pada penelitian tugas akhir ini telah dilakukan sistem monitoring laju aliran pada aliran saat melewati elbow tabs. Pada pengukuran laju aliran ini digunakan sensing elemen MPX2010DP dengan arduino uno sebagai mikrokontroller dan display monitoring dengan menggunakan visual studio 2013. Monitoring dilakukan dengan flowrate sebagai variable pengukuran. Berdasarkan hasil pengujian, didapatkan bahwa pada pengukuran flowrate diperoleh nilai ketidakpastian terbesar adalah ± 0,0533654 LPM, tingkat keakurasian sebesar 99,12 % dengan tingkat kepercayaan 95%.
========================================================================================================
Measurement of fluid flowrate is a measure to obtain data for analysis because in principle the fluid is pressurized energy that the more developed the processes in the industry. Many methods of measurement of flowrate used in fluid mechanics, any method used will produce a characteristic value that is different, not all types of measuring flowrates can use the same method, seen from the pressure drop generated will affect each of the methods used. In this research conducted flowrate measurements using tabs elbow. On the measurement of the fluid is often encountered as elbow pipe connection with a certain slope as 45 ° or 90 °. The slope on the elbow can affect the pressure difference and the pressure drop in the fluid passing through the elbow. In this research has been done flowrate monitoring system on the flow as it passes through tabs elbow. At this flowrate measurement used MPX2010DP sensing element with arduino uno as microcontroller and display monitoring using visual studio 2013. Monitoring is done with a variable flowrate measurement. Based on test results, it was found that the flowrate measurement values obtained greatest uncertainty is ± 0.0533654 LPM, the level of accuracy of 99.12% with a confidence level of 95%.

Item Type: Thesis (Diploma)
Additional Information: RSF 621.391 6 Suw r 3100017068513
Uncontrolled Keywords: Fluida, Pengukuran laju aliran, Elbow tabs, Monitoring
Subjects: T Technology > TJ Mechanical engineering and machinery > TJ213 Automatic control.
Divisions: Faculty of Industrial Technology > Physics Engineering > (D3) Diploma 3
Depositing User: Yeni Anita Gonti
Date Deposited: 03 Aug 2020 06:58
Last Modified: 04 Aug 2020 08:23
URI: http://repository.its.ac.id/id/eprint/76859

Actions (login required)

View Item View Item