Indrajaya, I Nyoman Rio (2021) Titrator Otomatis untuk Mengukur Kadar Kalsium Karbonat (CaCO3) pada Batu Kapur. Undergraduate thesis, Institut Teknologi Sepuluh Nopember.
Text
07111740000066-Undergraduate_Thesis.pdf - Accepted Version Restricted to Repository staff only until 1 October 2023. Download (2MB) | Request a copy |
Abstract
Kalsium Karbonat (CaCO3) merupakan mineral dominan yang ada pada batu kapur (limestone), umumnya kadarnya sekitar 95%. Kadar Kalsium Karbonat pada batu kapur dapat menjadi indikator dalam klasifikasi batu kapur yang nantinya akan diolah lagi dan digunakan dalam berbagai kebutuhan industri, pertanian, maupun kesehatan. Penentuan kadar kalsium karbonat dapat dilakukan melalui metode titrasi.
Titrasi merupakan salah satu metode analisis kuantitatif untuk menentukan konsentrasi dari suatu zat yang ada dalam larutan yang mengacu pada proses pengukuran volume titran yang diperlukan untuk mencapai titik ekuivalen. Metode ini dinilai cukup akurat sehingga metode titrasi konvensional sering dipakai hingga saat ini. Namun demikian, ada kelemahan pada metode titrasi konvensional, yaitu hasilnya sangat tergantung dari analis sehingga faktor human error dapat menurunkan akurasi dari pengukuran serta kapasitas kerja yang terbatas. Sebagai alternatif, titrator otomatis (autotitrator) dikembangkan untuk mengatasi keterbatasan tersebut.
Titrator otomatis pada Tugas Akhir ini menetukan hasil titrasi berdasarkan perubahan pH dan perubahan warna saat titik ekuivalen. Titran yang dialirkan oleh pompa peristaltik dikontrol oleh nilai pH larutan hingga sensor warna mendeteksi adanya perubahan warna larutan yang menandakan titik akhir titrasi. Berdasarkan hasil pengujian yang telah dilakukan, titrator otomatis ini dapat mengukur kadar Kalsium Karbonat pada Batu Kapur menggunakan metode titrasi dengan nilai error 4,2% untuk sampel dengan kadar 10,29%. Pada pengujian titrasi asam kuat dan basa kuat, titrator otomatis memiliki nilai error lebih besar karena perubahan nilai pH yang lebih cepat. Untuk penelitian selanjutnya, diperlukan sensor yang lebih responsif dan kontrol pompa peristaltik yang dapat menyesuaikan perubahan nilai pH pada titik ekuivalen.
==================================================================================================
Calcium Carbonate (CaCO3) is the dominant mineral in limestone, in general the level is around 95%. The level of calcium carbonate in limestone can be an indicator in the classification of limestone which will later be used again in various industrial, agricultural, and health needs. Calcium carbonate can be determined by titration method.
Titration is a quantitative analysis method for measuring the concentration of a substance in a solution which refers to the process of measuring the volume of the titrant needed to reach the equivalence point. This method is accurate enough that conventional methods are often used today. However, there are weaknesses in the conventional titration method, namely the results are very dependent on the analyst so that human error factors can reduce the measurement from the measurement and the limited working capacity. As an alternative, an automatic titrator (auto titrator) was developed to overcome this limitation.
The automatic titrator in this Final Project determines the titration results based on changes in pH and color changes at the equivalence point. The titrant flowed by the peristaltic pump is controlled by the pH value of the solution until the color sensor detects a change in the color of the solution which indicates the end point of the titration. Based on the results of the tests, this automatic titrator can measure the levels of Calcium Carbonate in Limestone using the titration method with an error value of 4.2% for samples with a concentration of 10.29%. In testing the titration of strong acids and strong bases, the automatic titrator has larger error value because the pH value changes faster. For further research, a more responsive sensor and peristaltic pump control are needed that can adjust changes in pH values at the equivalence point.
Item Type: | Thesis (Undergraduate) |
---|---|
Uncontrolled Keywords: | Larutan, CaCO3, Titrasi, Titrator Otomatis, Solution, CaCO3, Titration, Autotitrator |
Subjects: | T Technology > TK Electrical engineering. Electronics Nuclear engineering > TK7878 Electronic instruments |
Divisions: | Faculty of Intelligent Electrical and Informatics Technology (ELECTICS) > Electrical Engineering > 20201-(S1) Undergraduate Thesis |
Depositing User: | I Nyoman Rio Indrajaya |
Date Deposited: | 16 Aug 2021 04:17 |
Last Modified: | 16 Aug 2021 04:17 |
URI: | http://repository.its.ac.id/id/eprint/86934 |
Actions (login required)
View Item |