Rancang Bangun Sistem Akuisisi Citra Digital Untuk Keperluan Kuantifikasi Terbentuknya Corneal Arcus

Cahya, Deni Indra (2014) Rancang Bangun Sistem Akuisisi Citra Digital Untuk Keperluan Kuantifikasi Terbentuknya Corneal Arcus. Undergraduate thesis, Institut Teknologi Sepuluh Nopember Surabaya.

[thumbnail of 2410100081-Undergraduate_Thesis.pdf]
Preview
Text
2410100081-Undergraduate_Thesis.pdf - Published Version

Download (1MB) | Preview

Abstract

Mata merupakan indera penglihat pada manusia yang
terdiri dari beberapa bagian dengan karakteristik tertentu. Salah
satunya adalah kornea, yang dapat terlihat secara jelas dengan
kasat mata dibanding komponen lainnya. Pada kebanyakan
manusia berusia lanjut, sebuah cincin (ring) putih terbentuk
disekitar kornea yang menandakan adanya perubahan kadar lipid
dalam darah, cincin ini disebut corneal arcus. Cincin kornea ini
dapat digunakan sebagai identifikasi awal dari suatu penyakit.
Berapa penelitian awal telah membuktikan bahwa terbentuknya
corneal arcus ini berkolerasi positif dengan beberapa penyakit
seperti penyakit saluran pembuluh darah dan diabetes melitus.
Dalam penelitian tugas akhir ini dilakukan pengembangan sistem
metrologi optis untuk keperluan akuisisi citra digital guna
keperluan proses kuantifikasi lebih lanjut atas terbentuknya
corneal arcus. Sistem akusisi citra digital dirancang mengunakan
webcam, google welding sebagai housing sistem dan led 3 mm
untuk pencahayaan. Untuk memperoleh hasil kuantifikasi citra
yang akurat suatu proses kalibrasi dilakukan dengan berbagai pola
garis lurus dan lingkaran dengan diameter yang berbeda dengan
diketahui nilai ukuran sebenarnya. Dari hasil pengujian webcam
pada sistem akuisisi citra, letak mata pada sistem berkisar 4,607 ±
0,103 cm didepan webcam dengan persen relatif standar deviasi
sebesar 2,242%. Pada jarak rata-rata 4,6 cm didapat nilai
magnification sistem sebesar 0,105. Sistem akuisisi citra digital
memiliki akurasi yang baik dengan error tidak lebih dari 5%
berdasarkan hasil pengujian dengan mengukur jari-jari kornea sampel mata sehat
==================================================================================================
The human eye is an organ which is composed of several part with specific characteritics and function. One of those parts is Cornea. Cornea is surface of the human eye that can be seen clearly. Cornea can be used for identify sign of an illness. In older people’s cornea may form a circular white ring arround it, in medical term it is known as corneal arcus. Many studies mentioned that formation of corneal arcus is due to lipid depotistion in human blood. Some diseases are directly related to corneal arcus like diabetes melitus, and atherosclerosis. This research talk about development of optical metrology is performed by design of digital image acquisition system for the quantifying of cornea or formation of corneal arcus. This system may be used for study or further purpose research of development criteria diagnostic based on quantifying corneal arcus images.The system made of from webcam, google welding as housing and led for ligtning. To obtain a good result an acurrately quantify image this system is calibrated by many pattern like line and circular pattern. The data generated by the image acquisition system are uniform and calibrated to facilitate the process of identification and quantification. it was obtained that the average distance between the eye and the webcam is 4.607 ± 0.103 cm, with %RSD (relative standard deviation) of 2.242%. Around this average distance, the magnification will be M = 0.105. Using this magnification factor, the measurements
x
of test persons' corneal-ring diameter will produce very good accuracy, i.e. with error of 0 - 5 %

Item Type: Thesis (Undergraduate)
Additional Information: RSF 621.367 Cah r
Uncontrolled Keywords: Kornea mata, optical metrology, magnification, webcam
Subjects: T Technology > TA Engineering (General). Civil engineering (General) > TA1637 Image processing--Digital techniques
Divisions: Faculty of Industrial Technology > Physics Engineering > 30201-(S1) Undergraduate Thesis
Depositing User: EKO BUDI RAHARJO
Date Deposited: 07 Jan 2021 03:40
Last Modified: 07 Jan 2021 03:40
URI: http://repository.its.ac.id/id/eprint/82375