Implan tulang merupakan suatu alat medis yang terbuat dari logam yang ditanamkan ke dalam tubuh. Fungsi dari implan untuk menggantikan struktur dan fungsi tulang.Penelitian ini bertujuan untuk mengetahui laju korosi spesimen SS 316 L dan Baja ST 41 yang dilapisi Nickel Chrome (NiCr) dengan variasi perendaman 12 jam, 168 jam, 240 jam, dan 336 jam. SS 316 L merupakan material implan tulang yang sering digunakan saat ini. Jenis pengujian yang dilakukan pada penilitian ini antara lain Potentiodynamic Polarization (PDP), Electrochemical Impedance Spectroscopy (EIS), Scanning Electron Microscope (SEM), dan Scanning Electron Microscope Energy Dispersive X-Ray (SEM-EDX).Hasil pengujian menunjukan pola laju korosi yang sama pada kedua spesimen saat variasi perendaman selama 168 jam. Didapatkan nilai laju korosi tertinggi yaitu pada spesimen ST 41 yang dilapisi Nickel Chrome (NiCr) dengan nilai 0,024665 mm/year dan laju korosi terendah yaitu pada SS 316 L dengan nilai 0,0013166 mm/year pada variasi perendaman 336 jam.
====================================================================================================================================
A bone implant is a medical device made of metal that is implanted in the body. The function of the implant is to replace the structure and function of the bone.This study aims to determine the corrosion rate of SS 316 L specimens and ST 41 steel coated by Nickel Chrome (NiCr) with variations of immersion in 12 hours, 168 hours, 240 hours and 336 hours. SS 316 L is a bone implant material that is often used today. Types of tests conducted in this study include Potentiodynamic Polarization (PDP), Electrochemical Impedance Spectroscopy (EIS), Scanning Electron Microscope (SEM), and Scanning Electron Microscope Energy Dispersive X-Ray (SEM-EDX).The test results showed the same pattern of corrosion rate in both specimens when soaking variation for 168 hours. The highest corrosion rate was obtained in the ST 41 specimen coated with Nickel Chrome (NiCr) with a value of 0.024665 mm / year and the lowest corrosion rate was at SS 316 L with a value of 0.0013166 mm / year on a 336 hour immersion variation.
Item Type: | Thesis (Diploma) |
---|---|
Additional Information: | RSMI 620.112 23 Pam a-1 |
Uncontrolled Keywords: | Implan Tulang, Pelapisan, Nickel Chrome (NiCr), SS 316 L, Baja Karbon ST 41 |
Subjects: | T Technology > TJ Mechanical engineering and machinery |