Mikroskop merupakan alat yang digunakan untuk
melakukan pengamatan mikroorganisme. Salah satu kegiatan yang
dilakukan sebelum melakukan pengamatan mikroorganisme dengan
mikroskop digital adalah menempatkan objek pengamatan dalam
kaca preparat. Untuk mendapatkan hasil pengamatan yang terbaik
posisi preparat harus di atur sedemikian rupa sehinga mendapatkan
hasil yang diinginkan. Untuk saat ini sistem yang ada untuk
menentukan posisi preparat masih mengunakan cara konvensional,
sehinga masih memiliki kekurangan yaitu kurang presisi, proses
yang sulit dan memerlukan waktu yang cukup lama. Selain itu citra
yang dihasilkan oleh mikroskop digital hanya mencakup area yang
sempit, padahal citra yang dibutuhkan biasanya mencakup area yang
luas. Dengan demikian untuk mengatasi hal tersebut, diperlukan
perangkat yang dapat mengatur posisi preparat secara otomatis pada
mikroskop digital dan dapat membentuk citra panorama yang
mencakup area yang luas. Pada proses pengujian dan analisa
pengendali preparat mengunakan motor stepper didapatkan rata-rata
error gerakan preparat sebesar 5.6%. Pada pengujian proses stitching
untuk mendapatkan hasil maximum overlap citra diantara 50%
sampai 60%.
====================================================================================================
Microscope is a tool used to make observations of
microorganisms. One of the activities in microorganisms
observation with a digital microscope is placing objects in the glass
slide. To get the best result in microorganisms observation, glass
slide position should be set in such a way. For now existing system
to determine the position of glass slide still use conventional way, so
that have the disadvantage like less precision, difficult process and
requires a long time. In addition, the image produced by a digital
microscope cover only a small area, whereas the required images
usually cover a large area. Thus, to overcome this, we need a device
that can automatically adjust the position of glass slide in a digital
microscope and can form a panoramic image that covers a large
area. In the process of testing and analysis glass slide controller on
microscope digital using stepper motor resulting average error
movements approximately 5.6 %. In the process of testing and
analysis stitching technic to get the maximum result overlap
between the image 50% to 60%.

Item Type: Thesis (Undergraduate)
Additional Information: RSE 681.413 Huk p
Uncontrolled Keywords: Mikroskop Digital, Citra Panorama, Preparat
Subjects: T Technology > TA Engineering (General). Civil engineering (General) > TA1637 Image processing--Digital techniques
Divisions: Faculty of Industrial Technology > Electrical Engineering > 20201-(S1) Undergraduate Thesis
Depositing User: Mr. Tondo Indra Nyata
Date Deposited: 28 Jun 2018 03:12
Last Modified: 28 Jun 2018 03:12
URI: http://repository.its.ac.id/id/eprint/52055

Actions (login required)

View Item View Item