Optimasi Efek Radiobiologi Pada Organ Beresiko Dalam Perencanaan Radioterapi Dengan Teknik Vmat Untuk Kasus Kanker Orofaring

Hanum, Ulfi Qomariyah (2023) Optimasi Efek Radiobiologi Pada Organ Beresiko Dalam Perencanaan Radioterapi Dengan Teknik Vmat Untuk Kasus Kanker Orofaring. Masters thesis, Institut Teknologi Sepuluh Nopember.

[thumbnail of 6001221007-Master_Thesis.pdf] Text
6001221007-Master_Thesis.pdf - Accepted Version
Restricted to Repository staff only until 1 October 2025.

Download (3MB) | Request a copy

Abstract

Equivalent Uniform Dose untuk tumor adalah dosis ekuivalen biologis yang, jika diberikan secara seragam, menghasilkan pembunuhan sel yang sama dalam volume tumor dengan distribusi dosis sebenarnya yang tidak seragam. Umumnya digunakan dalam meminimalkan efek buruk radiobiologi. Penelitian bertujuan mengetahui nilai α generalized Equivalent Uniform Dose (gEUD) yang optimal untuk distribusi dosis pada masing-masing Organs At Risk (OARs) pada kasus kanker orofaring, evaluasi dosimetri dan radiobiologi dari masing-masing perencanaan radioterapi berdasarkan variasi nilai α gEUD, verifikasi antara distribusi dosis Treatment planning systems (TPS) dan pengukuran menggunakan dosimeter Film Gafchromic EBT3, Thermoluminescent Dosimeter (TLD) Chip, dan TLD Rod. Penelitian diawali dengan fabrikasi fantom serta anatomi pendukung yaitu jaringan kanker. Kemudian fantom dipindai dengan CT Scan untuk mendapatkan citra medis, kemudian dilakukan pembuatan perencanaan (TPS) dengan teknik Volumetric Modulated Arc Therapy (VMAT) 2 arcs. Perencanaan yang dibuat diberi variasi nilai α gEUD dengan rentang antara 0,1−40. Dosis yang diterima oleh masing-masing OARs yaitu otak, brainstem, mata kanan, mata kiri, parotis, dan spinal cord memiliki nilai yang sesuai dengan batas limitasi dosis yang telah ditentukan. Distribusi dosis pada Planning Target Volume (PTV) dengan parameter Homogeneity Index (HI) didapatkan nilai mendekati 0 sehingga dapat dikatakan homogen untuk nilai α gEUD 0,1 hingga 40. Probabilitas komplikasi jaringan normal mendekati 0% untuk semua OARs pada nilai α gEUD 0,1 hingga 5, sedangkan pada α gEUD di atas 10 menujukkan nilai Normal Tissue Complication Probability (NTCP) yang lebih tinggi. Probabilitas kontrol tumor bernilai mendekati 100% untuk nilai α gEUD 0,1 hingga 40. Verifikasi distribusi dosis dengan dosimeter film gafchromic EBT3 menunjukkan error kurang dari 10%. Verifikasi distribusi dosis dengan dosimeter TLD menunjukkan error antara 65 ̶ 74%.
================================================================================================================================
Equivalent Uniform Dose to the tumor is the biologically equivalent dose that, when administered uniformly, results in the same killing of cells within a tumor volume with a non-uniform distribution of the true dose. is a simple parameter designed to allow comparison between treatment plans when irradiation gain is not homogeneous. Commonly used in minimizing the adverse effects of radiobiology. This study aimed to determine the optimal α generalized Equivalent Uniform Dose (gEUD) value for the distribution of doses for each Organ At Risk (OARs) in cases of oropharyngeal cancer, dosimetric and radiobiological evaluation of each radiotherapy plan based on variations in α gEUD values, verification of dose distribution between Treatment Planning Systems (TPS) and measurements using a Gafchromic EBT3 Film dosimeter, Thermoluminescent Dosimeter (TLD) Chip, and TLD Rod. The research begins with phantom fabrication and supporting anatomy, namely cancer tissue. The phantom was scanned with a CT scan to obtain medical images, and planning (TPS) was then made using the Volumetric Modulated Arc Therapy (VMAT) 2 arcs technique. The plans made were given a variety of α gEUD values with a range between 0.1−40. The dose received by each OAR, namely the brain, brainstem, right eye, left eye, parotid, and spinal cord, has a value that corresponds to a predetermined dose limit. The distribution of doses on the Planning Target Volume (PTV) with the Homogeneity Index (HI) parameter obtained a value close to 0 so that it can be said to be homogeneous for α gEUD values of 0.1 to 40. The probability of normal tissue complications is close to 0% for all OARs at α gEUD values of 0.1 to 5, whereas α gEUD above 10 indicates a higher normal tissue complication probability (NTCP). The probability of tumor control was close to 100% for α gEUD values of 0.1 to 40. Verifying the dose distribution with a gafchromic EBT3 dosimeter showed an error of less than 10%. Verifying the dose distribution with a TLD dosimeter showed an error between 65 ̶ 74%.

Item Type: Thesis (Masters)
Uncontrolled Keywords: gEUD, Treatment Planning System, Verifikasi, Verification
Subjects: Q Science
Q Science > QC Physics
Divisions: Faculty of Science and Data Analytics (SCIENTICS) > Physics > 45101-(S2) Master Thesis
Depositing User: Ulfi Qomariyah Hanum
Date Deposited: 30 May 2023 05:05
Last Modified: 30 May 2023 05:05
URI: http://repository.its.ac.id/id/eprint/98020