Analisis ketebalan boks dan matrass inkubator bayi berkorelasi terhadap perubahan suhu

Setiawan, Singgih Yudha (2016) Analisis ketebalan boks dan matrass inkubator bayi berkorelasi terhadap perubahan suhu. Masters thesis, Institut Teknologi Sepuluh Nopember.

[thumbnail of 2414201004-master thesis.pdf]
Preview
Text
2414201004-master thesis.pdf

Download (1MB) | Preview

Abstract

Baby incubator adalah perangkat yang terdiri dari selungkup
seperti kotak kaku di mana bayi dapat disimpan dalam lingkungan yang
terkendali. Inkubator bayi sangat dibutuhkan oleh seorang bayi yang baru lahir,
baik bayi yang lahir dengan berat lahir rendah (BBLR) atau pun bayi yang lahir
normal yang kemungkinan mengalami hipotermia. dengan permasalahan
perbedaan selisih suhu didalam boks inkubator bayi perlunya dilakukan
penelitian ini, pada ketebalan bahan dari boks dan. suhu setting mulai dari 32 0 c,
340 c, 360 c, Ketebalan boks ( 5 mm dan 10 mm), ketebalan matras ( 5 cm, 10 cm,
15 cm) digunakan untuk pengambilan data yang akan ditangkap menggunakan
Incu analyzer. Uji Koefisien Korelasi Pearson pada Boks 5mm menunjukkan pada
ketebalan Matrass 5Cm mempunyai Koefisien Korelasi tertinggi sebesar 0,983
pada setting suhu 36oC. Ketebalan Matrass 10cm mempunyai koefisien korelasi
tertinggi 0,979 pada setting suhu 34oC. Ketebalan Matrass 15cm mempunyai
koefisien korelasi tertinggi 0,975 pada setting suhu 34oC. Uji Koefisien Korelasi
Pearson pada Boks 10mm menunjukkan pada ketebalan Matrass 5Cm mempunyai
Koefisien Korelasi tertinggi sebesar 0,973 pada setting suhu 36oC. Ketebalan
Matrass 10cm mempunyai koefisien korelasi tertinggi 0,978 pada setting suhu
36oC. Ketebalan Matrass 15cm mempunyai koefisien korelasi tertinggi 0,988 pada
setting suhu 36oC

==========================================================================================================

Baby incubator is a device consisting of a rigid box, like enclosure in
which the baby can be stored in a controlled environment. Incubator baby
urgently needed for a newborn, both babies born with low birth weight (LBW)
babies born or normal are likely to experience hypothermia. with the problem of
the difference the difference of temperature in the incubator baby crib need done
this research, on the thickness of the material and box. the temperature is settings
range from 32 0 c, 340 c, 360 c, the thickness of the cribs (5 mm and 10 mm),
thickness of the mattress (5 cm, 10 cm, 15 cm) is used for the retrieval of data that
will be captured using the Incu analyzer. Pearson Correlation Coefficient test on 5
mm thickness refer Box Matrass 5 cm has the highest Correlation Coefficient of
temperature settings on 0.983 360 c. Matrass thickness 10 cm has the highest
correlation coefficient on temperature setting 0.979 340 c. Matrass thickness 15
cm has the highest correlation coefficient on temperature setting 0.975 340 c.
Pearson Correlation Coefficient tests on 10 mm thickness refer Box Matrass
highest Correlation Coefficient has a 5 cm of 0.973 on the setting temperature of
360 c. Matrass thickness 10 cm has the highest correlation coefficient 0.978 on
setting temperature of 360 c. Matrass thickness 15 cm has the highest correlation
coefficient 0.988 on setting temperature of 36oC

Item Type: Thesis (Masters)
Additional Information: RTF 620.112 15 Set a
Uncontrolled Keywords: Ketebalan matrass,Ketebalan books, korelasi perubahan suhu
Subjects: Q Science > QC Physics > QC271 Temperature measurements
Divisions: Faculty of Industrial Technology > Physics Engineering > 30101-(S2) Master Thesis
Depositing User: - Taufiq Rahmanu
Date Deposited: 02 Dec 2019 06:15
Last Modified: 02 Dec 2019 06:15
URI: http://repository.its.ac.id/id/eprint/72145

Actions (login required)

View Item View Item